Dr. Goos Suprema GmbH
Lead Vest Triplex
Lead Vest Triplex
The Triplex lead vest (X-ray protective vest) offers comprehensive X-ray protection with optimum front and back protection. Its design ensures even weight distribution and increased wearing comfort through integrated back support. The Triplex vest offers the highest wearing comfort currently available.
- Excellent front and back protection
- Even weight distribution
- Integrated back support
- Ideal for long procedures
- Including chest pocket
Additional options
- iPron
- Thyroid protection
- Measure-to-fit service
- Standard sizes
- Embroidery
Lead equivalent
Lead equivalent
Protective materials
Protective materials
Protective effect of our protective materials
| Protective material | Lead equivalent | Area weight | 50 kV | 70 kV | 90 kV | 110 kV | 150 kV |
|---|---|---|---|---|---|---|---|
| SupraNorm (standard lead) | 0.25 mm Pb | 3.2–3.6 kg/m² | 99 % | 95 % | 90 % | 86 % | 84 % |
| SupraNorm (standard lead) | 0.35 mm Pb | 4.4–5.1 kg/m² | 99 % | 98 % | 94 % | 91 % | 86 % |
| SupraNorm (standard lead) | 0.50 mm Pb | 6.4–7.1 kg/m² | 99 % | 99 % | 96 % | 95 % | 92 % |
| SupraNorm (standard lead) | 1.00 mm Pb | 12.8–14.2 kg/m² | 99 % | 98 % | 97 % | 96 % | 95 % |
| SupraLight Plus (light lead) | 0.25 mm Pb | 3.2–3.4 kg/m² | 99 % | 96 % | 92 % | 89 % | 83 % |
| SupraLight Plus (light lead) | 0.35 mm Pb | 4.5–4.7 kg/m² | 99 % | 98 % | 95 % | 93 % | 90 % |
| SupraLight Plus (light lead) | 0.50 mm Pb | 6.4–6.7 kg/m² | 99 % | 99 % | 97 % | 96 % | 93 % |
| SupraSine (lead-free) | 0.25 mm Pb | 3.0–3.2 kg/m² | 99 % | 95 % | 91 % | 86 % | – |
| SupraSine (lead-free) | 0.35 mm Pb | 4.1–4.3 kg/m² | 99 % | 97 % | 94 % | 90 % | – |
| SupraSine (lead-free) | 0.50 mm Pb | 6.0–6.1 kg/m² | 99 % | 99 % | 97 % | 95 % | – |
The protective effect at the stated X-ray tube voltage corresponds to the requirements of EN 61331-1:2014. SupraNorm and SupraLight Plus: 50 to 150 kV, SupraSine: 50 to 110 kV (no values stated at 150 kV).
Cleaning instructions
Cleaning instructions
Cleaning & care of external materials “SupraTex”, “SupraSoft”, “SupraSoft Plus” and “SupraRib”.
Fresh soiling, in particular contrast agents, blood or other liquids, should be removed immediately with a lukewarm detergent solution and a soft cloth to prevent them from drying out. No aggressive cleaning agents containing alcohol (thinner, white spirit, spirits etc.) should be used. We recommend our cleaning agent: “SupraPflege” (Descosept-Spezial listed by VAH/ÖGHMP/IHO) for cleaning, care and disinfection of X-ray protective clothing.